Class II Biosafety cabinet – Type B1

  • Aeromech™ Class II Type B1 biological safety cabinet has a common plenum from which 70% of air is exhausted, and 30% re-circulated to the work area as the downflow.
  • Type B1 cabinets also have a dedicated exhaust feature that eliminates re-circulation when work is performed towards the back within the interior of the cabinet. Toxic chemicals employed as an adjunct to microbiological processes should only be employed if they do not interfere with work when re-circulated in the downflow.
1 Product Class II  Bio safety Cabinet –Type B1
2 Air balancing 30% recirculation, 70% exhaust
3 Protection  Product & Personnel & Environmental protection
4 Bio safety Level Bio safety Level 1 , 2 & 3
5 Exhaust Duct Requirement Hard Duct Only
6 Control Plenum surrounded by Negative Plenum
7 Cleanliness level ISO class 5 as per ISO 14644-1
8 Average Airflow Inflow Velocity -  Minimum 0.45 m/s ( 90 FPM) at 8” front opening
Velocity Down Velocity -  0.30 m/s ( 65 FPM)
9 Noise level 60 db on A scale
10 Light Intensity Minimum 900 lux inside the working area at ambient zero.
11 MOC Main Body -  Galvanized Iron with PU coated / Stainless Steel – Satin Finish
Working Area -  Stainless Steel IS 304 Grade Satin Finish
12 Supply Filter HEPA Filter having EU 13 rating with suitable size.
Efficiency 99.99% down to 0.3 Microns.
13 Exhaust Filter HEPA Filter having EU 13 rating with suitable size.
Efficiency 99.99% down to 0.3 Microns.
14 Pre Filter Aeromech® make, EU 4 rating with suitable size.
Efficiency 90% down to 10 Microns.
15 Noise Level Less than 70 db 'A'.
16 Vibration Level Minimum.
Instrumentation
17  Differential Pressure Gauge Magnehelic Gauge - 01 No. (make : Dwyer, USA) with each unit
(To measure differential pressure across the supply HEPA filter).
Motor Blower Assembly Statically & Dynamically balanced with suitable rating and size.
18  Supply & Exhaust (Induction motor with centrifugal blower) A Mounting unique spring suspension system to take care of vibration and noise.
19 3 Pins Switch & Socket 15 AMPS - 1 set in each unit.
20  Other Switches On/Off switches for motor blower & CFL lights.
21 Microprocessor Control Airflow display and UV Interlocking and Low airflow Alarm and Supply  & Exhaust filter Fail